題名: Computational comparison of different plating strategies in medial open-wedge high tibial osteotomy with lateral hinge fractures
作者: Chen, Yen-Nien;Chuang, Chang-Han;Yang, Tai-Hua;Chang, Chih-Wei;Li, Chun-Ting;Chang, Chia-Jung;Chang, Chih-Han
貢獻者: 高齡福祉科技研究所
關鍵詞: Finite element simulation;High tibial osteotomy;Lateral hinge fracture;Plating strategy
日期: 2020-09-01
上傳時間: 2026-10-27 17:11:21 (UTC+8)
摘要: Background: Lateral hinge fracture (LHF) is associated with nonunion and plate breakage in high tibial osteotomy (HTO). Mechanical studies investigating fixation strategies for LHFs to restore stability and avoid plate breakage are absent. This study used computer simulation to compare mechanical stabilities in HTO for different LHFs fixed with medial and bilateral locking plates.
Methods: A finite element knee model was created with HTO and three types of LHF, namely T1, T2, and T3 fractures, based on the Takeuchi classification. Either medial plating or bilateral plating was used to fix the HTO with LHFs. Furthermore, the significance of the locking screw at the combi hole (D-hole) of the medial TomoFix plate was evaluated.
Results: The osteotomy gap shortening distance increased from 0.53 to 0.76, 0.79, and 0.72 mm after T1, T2, and T3 LHFs, respectively, with medial plating only. Bilateral plating could efficiently restore stability and maintain the osteotomy gap. Furthermore, using the D-hole screw reduced the peak stress on the medial plate by 28.7% (from 495 to 353 MPa), 26.6% (from 470 to 345 MPa), and 32.6% (from 454 to 306 MPa) in T1, T2, and T3 LHFs, respectively.
Conclusion: Bilateral plating is a recommended strategy to restore HTO stability in LHFs. Furthermore, using a D-hole locking screw is strongly recommended to reduce the stress on the medial plate for lowering plate breakage risk.
關聯: Journal of Orthopaedic Surgery and Research, 15(1), e409. https://doi.org/10.1186/s13018-020-01922-0